MedEx Easy Breath Mesh Nebuliser

MedEx Easy Breath Mesh Nebuliser Caution For Usage

Manufacturer:

Y.S.P. Industries

Distributor:

Y.S.P. Industries
Full Prescribing Info
Caution For Usage
Do not attempt to clean mesh with any foreign objects, it may damage the mesh.
Avoid dropping the device otherwise it may not function normally.
Keep the device away from direct sunlight, excessive heat or cold to avoid damaging the batteries.
Do not attempt to open repair or modify this device.
Do not drop any liquids on the main unit or on the adapter, if you drop liquid on it, wipe it off immediately.
Follow local laws and recycling regarding disposal or recycling of components, batteries and packaging.
Cleaning and Disinfection: If the device is not cleaned and disinfected correctly and frequently as indicated, microorganisms may remain in the unit and increase risk of infection. After every use: Clean medication cup with distilled water after each inhalation.
1. Pour out the residual solutions in the medication cup.
2. Pour some distilled water into the medication cup.
3. Turn on the device to nebulize the distilled water for 1 to 2 minutes to clean the mesh.
4. Remove the batteries or disconnect the adapter from the main unit.
5. Remove the medication cup from the main unit.
6. Wash and rinse the medication cup with distilled water.
7. Shake off excess water and allow parts to be full air dried on a clean, dry towel.
8. Use gauze or clean towel to wipe off stains on the main unit if necessary.
9. Make sure that all cleaned parts are completely dry before storing them or use them next.
Keep the battery compartment dry all the time.
DO NOT poke the mesh with finger, cotton swab or any objects.
DO NOT clean parts in a dishwasher.
DO NOT use microwave to dry any parts.
It is important to disinfect the medication cup on a daily basis.
1. Disinfection by boiled water. Rinse the medication cup with distilled water.
Bring a saucepan of DISTILLED water to boil. Note: DO NOT boil medication cup directly.
Carefully immerse the medication cup in the boiled for a maximum of ten (10) minutes.
Carefully remove the medication cup from the boiled water and shake off excess water.
Allow parts to cool down and fully air dried on a clean, dry towel, and out of reach of children.
Make sure all cleaned parts are completely dry before storing them or use them next time.
2. Disinfection by alcohol: Rinse the medication cup with distilled water.
Immerse the medication cup in 75% ethyl alcohol for 1 minute.
Rinse the medication cup with distilled water again, shake off excess water and allow parts to be fully air dried on a clean, dry towel.
Make sure all cleaned parts are completely dry before storing them or use them next time.
The following items are able to be able to be disinfected by boiled water or alcohol: Medication cup (with cap).
Parts as follows are not able to be disinfected by boiled water or alcohol: Main unit; Power adapter.
DO NOT rinse or immerse the main unit in any liquid.